Technologies is involved with a number of instruments using microwave /
millimeter wave techniques for the characterization of physical properties
of various materials. The
instrument presented here is an example of a resonant cavity near field
microwave system. Here,
the energy from a resonant cavity is coupled outside the cavity via a
small hole or a conductive wire, producing evanescent fields. The size of
the hole or the wire is usually much smaller than the wavelength.
The hole or wire size (not the wavelength) sets the spatial resolution for
the imaging of materials. Depending upon the probe configuration, a
spatial resolution on the order of l/1000 (where l
is the resonant wavelength) can be achieved.
The important parameters of the resonator are the resonant
frequency (fo) and the quality factor (Q). A sample coupled
into a resonator can change fo and / or Q. A change in Q is
associated with surface resistance and dielectric losses. A shift in fo
is caused by the material’s dielectric constant. Probes of this type are
also called evanescent microwave probes (EMP).
The main probe features include the following: -
high sensitivity to material characteristics -
high spatial resolution - fraction of a wavelength -
sensitive to surface properties of metals - surface resistance -
sensitive to structural and elctromagnetic properties of semiconducting
materials and dielectric materials, including ceramics and composites
Figure 1. Shows the output from a scan over a 2.5 mm thick glass fiber composite plate, shown in the inset photograph. The plate contains 6 through holes of decreasing diameter going from left to right. For metallic samples, surface defects and variations, perhaps indicating corrosion, can be revealed. For ceramic and composite samples, such structural defects as voids, delaminations or disbonds, porosity, and corrosion, produce local impedance changes, revealing their presence. A near field microwave instrument is only sensitive to surface and near-surface defects on metal samples; for dielectric samples, the thickness of the measured structures can be on the order of several centimeters, depending on the microwave wavelength and the scanning configuration. For
the data presented here, the sample under test was directly coupled to a 9
GHz high-Q resonant cavity through a several millimeter sized rectangular
aperture, thus providing sub-wavelength resolution. In the microwave
circuit, the resonant cavity was coupled so as to nearly match the
impedance of the input circuit, so that only a minimal amount of power was
reflected back. An advanced frequency modulation and feedback circuit,
implemented in the present system, varied the frequency to lock the
circuit on to the resonant frequency defined by the resonant cavity /
sample combination.
Figure 2. In this figure, the sample is a 6.4 mm thick, square, glass
fiber composite plate. There
are 9 drill markings on this plate, going from through holes to just
shallow indentations. The
photo in the lower right shows the plate from the back side, meaning that
only the through holes mar the surface; some of the other markings are
visible simply because the plate thickness is thinner below the markings.
The markings are in a 3 x 3 array; note that there is no visible
sign of the marking at the lower left.
If the plate is turned over, this lower left marking is as shown
(see the photo in the lower left); this marking is cone shaped with a 2.3
mm maximum depth and 9 mm diameter at the surface of the plate ¾
recall that the plate is 6.4 mm thick, so this marking is over 4 mm below
the other surface of the plate. The
output shown was produced when scanning the plate from the backside, over
the marking shown in the lower left photo.
If this marking represented a structural defect, its presence would
be clearly revealed. The
variations in output shown in these figures correspond to changes in the
reflected power — while using the feedback circuit to lock on to the
resonant frequency of the cavity / sample grouping — as the aperture of
the cavity is scanned over the sample. The changes in reflected power
correspond to changes in the complex impedance of the cavity / sample
combination; these impedance changes in turn reveal properties of the
portion of sample directly below the aperture.
Figure 3: This is another example showing changes in reflected power caused by markings on a 2.3 mm thick fiberglass plate. The upper portion of the figure is a photo of the plate — moving from the left to the right side: 0.77 mm diameter blind hole; 0.77 mm diameter through hole; 0.61 mm diameter blind hole; 0.49 mm diameter blind hole (the blind holes extend approximately halfway through the plate.
4a 4b
4c
4d
4e
4f
Figure 4: Scan of various markings in a 5 mm thick SiC ceramic plate: (a) is a photograph of the whole plate and (b) shows just the region of the 3 markings. Marking N1, in (b), is an annular groove, 4 mm OD X 2 mm ID X 2 mm deep. (c) shows the output of a scan directly over N1, clearly revealing this feature in a ceramic plate. (d) shows a scan of the same marking from the back side; this surface. In (b), N2 is a 1.5 mm diameter counterbore, less than 0.5 mm deep; a scan of this feature is shown in (e). N3 in (b) is 0.75 mm counterbore, less than 0.5 mm deep; a scan of this feature is shown in (f). For
non-destructive inspection / evaluation of dielectric samples, this near
field microwave instrument has definite advantages over the other
currently used techniques. Mechanical
testing: frequently
destructive, only gives indirect information regarding the investigated
structure. Electron
scanning microscopy: limited
to surface characterization. Conventional
ultrasound: requires fluid
coupling to the surface, which is time consuming and inconvenient. X-ray
characterization: not user
friendly in the sense that the measurement must be shielded from the
operator. Thermal
techniques: data
interpretation difficult and confusing, especially for thick samples. We
feel this microwave near field, high-Q tool is a powerful NDE / NDI
instrument. Design variations allow focusing on different applications.
Tristan can provide a microwave near field scanner — including sensor
(consisting of a resonant cavity at a specified frequency with various E
and H field orientations outside the cavity), scan stage, computer control
and data analysis ¾
for use in Non-Destructive Evaluation
and Non-Destructive Inspection (NDE & NDI) applications, as
well as physical properties characterization. Contact
us concerning your specific needs.
|